01 4503452
EATON CORPORATION
ISOLATION TRANSFORMER 115/230V MED GRADE C14 INLET 4 C13 OUT

ISOLATION TRANSFORMER 115/230V MED GRADE C14 INLET 4 C13 OUT

Part Number:
IS300HGDV

Introduction

The IS300HGDV Isolator® Series Dual-Voltage 115/230V 300W 50/60Hz Medical-Grade™ Transformer with full CAN/ES/IEC/EN 60601-1 listing is ideal for protecting sensitive electronic equipment in patient-care vicinities. It can be configured for either step-up voltage, step-down voltage or pass-through voltage operation. Along with line isolation and continuous EMI/RFI line noise filtering, its surge protection rating of 240 joules suppresses excess voltage to prevent it from damaging connected equipment.

An internal low-impedance isolation transformer component offers 100% isolation from the input mains AC line. Faraday shield reduces the cumulative leakage current of the IS300HGDV and connected equipment to levels less than 50 microamps, while removing utility switching transients and harmonics generated by other on-site loads and reducing the impact of utility-related surge conditions.

Transformer filtering provides continuous common-mode noise rejection with no wearable parts. The IS300HGDV reduces 6000V IEEE 587 Category A and B ring wave and combination wave test surges to only 0.5V common mode.

Components connect to four C13 outlets with fuse overload protection and a lighted on/off switch. The rugged all-metal unit housing ships ready for placement in any industrial, medical, office, network or audio/video environment. Requires a C13 AC input cord with country specific plug (sold separately).

Full CAN/ES/IEC/EN 60601-1 Medical-Grade Listing with C14 Inlet and 4 C13 Outlets
Ideal for protecting sensitive electronic equipment in patient-care vicinities
Reduces cumulative leakage current of Isolator and connected equipment to levels less than 50 µA
Inexpensive alternative to dedicated circuits and site electrical upgrades
Rugged all-metal housing designed to withstand frequent use in busy environments
Requires a C13 AC input cord with country specific plug (sold separately)

Selectable 115/230V Input/Output Voltage Suitable for Most Applications
Step-up, step-down or pass-through voltage operation (input frequency always equals output frequency)
Internal low-impedance toroidal isolation transformer with Faraday shield offers 100% isolation from mains AC power line
Automatically shuts down all outlets in case of overload
Ships ready for placement in any industrial, medical, office, network or audio/video environment

Protects Connected Equipment from Dangerous Spikes, Surges and Line Noise
Line isolation, continuous noise filtering and 240 joules of AC surge protection
Removes EMI/RFI noise, utility switching transients, load-generated harmonics and ground loops
Floating AC output prevents noise coupling from noisy ground circuits
Reduces 6000V IEEE 587 Category A and B ring wave and combination wave test surges to only 0.5V common mode

Features

AC inlet type

C20 coupler

AC outlet types

C13

AC outlets quantity

4 AC outlet(s)

AC input voltage

115 - 230 V

AC input frequency

50 - 60 Hz

Input current (max)

3.15 A

Short circuit protection

Y

Load capacity

300 VA

Housing colour

White

Certification

ANSI/AAMI ES60601-1:2005+A2+A1, IEC60601-1 3RD EDITION +A1, cAN/CSA-C22.2 NO. 60601-1:14, CE, NOM

Operational conditions

Operating temperature (T-T)

0 - 40 °C

Storage temperature (T-T)

-15 - 50 °C

Operating relative humidity (H-H)

5 - 95%

Operating altitude

0 - 3000 m

Weight & dimensions

Width

150.1 mm

Depth

199.9 mm

Height

110 mm

Weight

5.6 kg

Package width

249.9 mm

Package depth

325.1 mm

Package height

210.1 mm

Package weight

6.18 kg

Logistics data

Products per master (outer) case

1 pc(s)

Master (outer) case width

249.9 mm

Master (outer) case length

325.1 mm

Master (outer) case height

210.1 mm

Master (outer) case gross weight

6.18 kg

Country of origin

China

No options for this product
.